SX2 series muffle furnace can repair the defects such as elongation, necking, thickness reduction, section deformation, bending and bulging caused by high temperature creep. To this end, the muffle furnace tube is first fully measured prior to repair to determine the repair process and the acceptance criteria after repair. Skin Care Product,Led Light Therapy,Blackhead Remover Tool,Silicone Facial Cleanser Brush Shenzhen Jie Zhong Lian Investment Co., Ltd. , https://www.meizons.com Troubleshoot the reason why the muffle furnace cycle does not heat up
Next, the defect is removed and then welded. Stainless steel muffle furnace product analysis and technical improvement, stainless steel with bright annealing large production line, its total height is 55m, can handle stainless steel strip with width of 600 ~ 1250mm, thickness of 0.1 ~ 3.0mm, the important accessory is muffle furnace Handle stainless steel.
Muffle furnace cycle operation type, for laboratory, industrial and mining enterprises, scientific research units for elemental analysis and determination of general small steel parts quenching, annealing, tempering and other heat treatment, high temperature Mafu furnace can also be used for metal, ceramic sintering, dissolution , analysis and other high temperature heating.
How to check if the muffle does not heat up?
1. After the muffle furnace is naturally cooled to the normal temperature state, close the main power supply, turn on the temperature control switch, and observe whether the temperature meter display is normal. Red light, green light conversion is normal, if everything is normal, it means there is no problem with the temperature controller, otherwise it is still.
2. Turn off the main power supply and temperature control switch, use a multimeter to detect the terminal behind the muffle furnace, measure whether the resistance value is normal, the resistance is infinitely large, indicating that the open circuit, the electric furnace wire is blown, and the resistance deviates from the rated value, indicating that the electric furnace wire is seriously oxidized.
3. Detect whether the various terminals are fully linked, and open the controller to measure whether the contacts of the AC contactor can be reliably sucked.